Atec Rohde-Schwarz-ZVA Series User Manual

Page 32

Advertising
background image

Version 10.00, May 2012

32

Rohde & Schwarz

R&S

®

ZVA Vector Network Analyzer

Receiver step attenuators

These attenuators permit the input signal level to be attenuated
in 5 dB steps up to 35 dB.

Frequency range

R&S

®

ZVA8

300 kHz to 8 GHz

R&S

®

ZVA24

10 MHz to 24 GHz

R&S

®

ZVA40

10 MHz to 40 GHz

R&S

®

ZVA50

10 MHz to 50 GHz

R&S

®

ZVA67

10 MHz to 67 GHz

Attenuation

0 dB to 35 dB

Attenuation steps

5 dB

Attenuation accuracy

< 2 dB

Dynamic range

R&S

®

ZVA8

300 kHz to 8 GHz

is reduced by 1 dB

R&S

®

ZVA24

10 MHz to 13 GHz

is reduced by 1 dB

13 GHz to 24 GHz

is reduced by 2 dB

R&S

®

ZVA40

10 MHz to 13 GHz

is reduced by 1 dB

13 GHz to 24 GHz

is reduced by 2 dB

24 GHz to 40 GHz

is reduced by 3 dB

R&S

®

ZVA50

10 MHz to 13 GHz

is reduced by 1 dB

13 GHz to 24 GHz

is reduced by 2 dB

24 GHz to 50 GHz

is reduced by 3 dB

R&S

®

ZVA67

10 MHz to 13 GHz

is reduced by 1 dB

13 GHz to 24 GHz

is reduced by 2 dB

24 GHz to 67 GHz

is reduced by 3 dB

Noise level

R&S

®

ZVA8

300 kHz to 8 GHz

is increased by 1 dB

R&S

®

ZVA24

10 MHz to 13 GHz

is increased by 1 dB

13 GHz to 24 GHz

is increased by 2 dB

R&S

®

ZVA40

10 MHz to 13 GHz

is increased by 1 dB

13 GHz to 24 GHz

is increased by 2 dB

24 GHz to 40 GHz

is increased by 3 dB

R&S

®

ZVA50

10 MHz to 13 GHz

is increased by 1 dB

13 GHz to 24 GHz

is increased by 2 dB

24 GHz to 50 GHz

is increased by 3 dB

R&S

®

ZVA67

10 MHz to 13 GHz

is increased by 1 dB

13 GHz to 24 GHz

is increased by 2 dB

24 GHz to 67 GHz

is increased by 3 dB

Advertising