Calibration limits, Cell constant, Zero calibration procedure – Hach-Lange POLYMETRON 9500 Conductivity Module User Manual User Manual

Page 13

Advertising
background image

Contacting conductivity sensors: Use air (zero calibration) and the process sample to define the
calibration curve. When the process sample is used, the reference value must be determined with a
certified secondary verification instrument.
Inductive conductivity sensors: Use air (zero calibration) and a reference solution or process
sample to define the calibration curve. A reference solution calibration is recommended for best
accuracy. When the process sample is used, the reference value must be determined with a certified
secondary verification instrument.

Calibration limits

A calibration will fail if the value is ± 10% of the target value.

Cell constant

Before making a calibration make sure the sensor cell parameters are correct.

1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.
3. Select CELL CONSTANT and push enter.
4. Contacting conductivity sensors: Select the cell K range for the sensor (0.01, 0.1 or 1.0) then

enter the actual K value as printed on the label attached to the sensor.
Inductive conductivity sensors: Enter the actual K value as printed on the label attached to the
sensor.

Zero calibration procedure

Use the zero calibration procedure to define the unique zero point of the sensor.

1. Remove the sensor from the process. Wipe with a clean towel to make sure the sensor is dry.
2. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
3. If the pass code is enabled in the security menu for the controller, enter the pass code.
4. Select ZERO CAL and push enter.
5. Select the option for the output signal during calibration:

Option

Description

ACTIVE

The instrument sends the current measured output value during the calibration procedure.

HOLD

The sensor output value is held at the current measured value during the calibration procedure.

TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change

the preset value.

6. Place the sensor in air, push enter.
7. Review the calibration result:

• PASS—the sensor is calibrated and ready to measure samples.
• FAIL—the calibration is outside of accepted limits. Clean the sensor and retry. Refer to

Troubleshooting

on page 18 for more information.

8. If the calibration passed, push enter to continue.
9. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer

to

Change calibration options

on page 17.

10. On the NEW SENSOR screen, select whether the sensor is new:

Option Description

YES

The sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.

NO

The sensor was calibrated previously with this controller.

English 13

Advertising