Hach-Lange POLYMETRON 9500 Conductivity Module User Manual User Manual

Page 6

Advertising
background image

Note: If the sensor cable is not long enough to reach the controller, an interconnect cable (model 150727,10000)
and junction box (model 08335=A=6000) should be used to extend the distance.

Table 1 Conductivity sensor wiring

Terminal

Description

Sensor

8398

831x and 8394

1

Inner electrode

White (white cable)

White

2

Inner electrode

Black (white cable)

—

3

Ground

—

Green

4

—

—

—

5

—

—

—

6

—

—

—

7

Temp -

Green

Black

8

—

—

—

9

—

—

—

10

Temp +

Yellow

Yellow

11

Outer electrode

Brown (brown cable)

Red

12

Outer electrode

Black (brown cable)

—

In Step 3 of the following sequence, set the switch as follows:

Switch Setting

Description

Setting 1

Contacting conductivity sensor

Setting 2

Inductive conductivity sensor

6 English

Advertising