Temperature calibration, Exit calibration procedure, Change calibration options – Hach-Lange POLYMETRON 9500 Conductivity Module User Manual User Manual

Page 17

Advertising
background image

7. If the calibration passed, push enter to continue.
8. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer

to

Change calibration options

on page 17.

9. On the NEW SENSOR screen, select whether the sensor is new:

Option Description

YES

The sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.

NO

The sensor was calibrated previously with this controller.

10. Reconnect the sensor, return it to the process and push enter. The output signal returns to the

active state and the measured sample value is shown on the measure screen.

Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.

Temperature calibration

The temperature sensor has been calibrated at the factory. However, it is recommended to always
calibrate the temperature sensor before calibrating the measurement sensor.

1. Put the sensor in a container of water that is at a known temperature. Measure the temperature of

the water with an accurate thermometer or independent instrument.

2. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
3. If the pass code is enabled in the security menu for the controller, enter the pass code.
4. Select 1 PT TEMP CAL and push enter.
5. The raw temperature value is displayed. Push enter.
6. Enter the correct value if different from that displayed and push enter.
7. Push enter to confirm the calibration.

Exit calibration procedure

If the back key is pushed during a calibration, the user can exit the calibration.

1. Push the back key during a calibration. Three options are shown:

Option

Description

QUIT CAL

Stop the calibration. A new calibration must start from the beginning.

BACK TO CAL Return to the calibration.

LEAVE CAL

Exit the calibration temporarily. Access to other menus is allowed. To return to the
calibration, push the menu key and select SENSOR SETUP.

2. Use the arrow keys to select one of the options and push enter.

Change calibration options

The user can set a calibration reminder or include an operator ID with calibration data from this
menu.

1. Push the menu key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.

English 17