Table 2 Sensor DIAG/TEST menu (continued)

Option

Description

SENSOR SIGNALS

Shows the current sensor signal information.

SENSOR DAYS

Shows the number of days that the sensor has been in
operation.

RESET SENSOR

Resets the number of days that the sensor has been in
operation and resets all calibration data to defaults.

Error list

Errors may occur for various reasons. An error icon consists of an exclamation point within a circle.
When an error occurs, the error icon and the measurement screen flash alternately in the main
display. All outputs are held when specified in the controller menu. To view errors, push the menu
key and select DIAGNOSTICS. Then select the device to view any problems associated with that
device.
A list of possible errors is shown in

Table 3

.

Table 3 Error list for conductivity sensors

Error

Description

Resolution

ADC FAILURE

The analog to digital conversion
failed

Make sure that the sensor module is
fully inserted into the controller
connector. Replace the sensor
module.

SENSOR MISSING

The sensor is missing or
disconnected

Examine the wiring and connections
for the sensor and for the module.
Make sure that the terminal block is
fully inserted into the module.

SENS OUT RANGE

The sensor signal is outside of the
accepted limits (2 S/cm)

Make sure that the display format is
set for the correct measurement
range.

Warning list

A warning icon consists of an exclamation point within a triangle. Warning icons appear on the right
of the main display below the measurement value. A warning does not affect the operation of menus,
relays and outputs. To view warnings, push the menu key and select DIAGNOSTICS. Then select
the device to view any problems associated with that device. The warning icon will no longer be
displayed once the problem has been corrected or acknowledged.
A list of possible warnings is shown in

Table 4

.

Table 4 Warning list for conductivity sensors

Warning

Description

Resolution

MEAS TOO HIGH

The measured value is > 2 S/cm,
1,000,000 ppm, 200% or 20,000 ppt

Make sure that the display format is
set for the correct measurement
range

MEAS TOO LOW

The measured value is < 0 μS/cm,
0 ppm, 0% or 0 ppt

Make sure that the sensor is
configured for the correct cell
constant.

ZERO TOO HIGH

The zero calibration value is too
high

Make sure that the sensor is held in
air during zero calibration and is not
located near radio frequency or
electromagnetic interference. Make
sure that the cable is shielded by
metal conduit.

ZERO TOO LOW

The zero calibration value is too low

20 English