User manual, Specifications, Contacting conductivity sensors – Hach-Lange POLYMETRON 9500 Conductivity Module User Manual User Manual

Page 3: English

Advertising
background image

Specifications

Specifications are subject to change without notice.

Contacting conductivity sensors

Specification

8310

8311

8312

Application

Drinking water and wastewater treatment

Chemical processes

Demineralized and softened water

Body material

Black PSU

Black PSU

Black PSU

Internal electrode

SS 316L

SS 316L

Graphite

External electrode

SS 316L

SS 316L

Graphite

Insulator

PSU

PSU

PSU

Connector

Glass polyester (IP65)

Glass polyester (IP65)

Glass polyester (IP65)

Cell constant K

0.01 (cm

-1

)

0.1 (cm

-1

)

1.0 (cm

-1

)

Accuracy

< 2%

< 2%

< 2%

Measurement range

0.01—200 μS.cm

-1

0.1 μS—2 mS.cm

-1

1 μS—20 mS.cm

-1

Temperature response

< 30 seconds

< 30 seconds

< 30 seconds

Maximum temperature

125 °C

125 °C

125 °C

Maximum pressure

10 bars

10 bars

10 bars

Sample connection

¾ inch NPT

¾ inch NPT

¾ inch NPT

Specification

8315

8316

8317

8394

Application

Pure water production

Process water monitoring (condensates, cleaning cycles, heat

exchangers)

Monitoring of

ultrapure water in

pharmaceutical and

food industries

Suitable for CIP-SIP

processes

Body material

SS 316L

SS 316L

SS 316L

SS 316L

Internal electrode

SS 316L

SS 316L

Graphite

SS 316L

External electrode

SS 316L

SS 316L

Graphite

SS 316L

Insulator

PES

PES

PES

PEEK

Connector

Glass polyester

(IP65)

Glass polyester

(IP65)

Glass polyester

(IP65)

Glass polyester

(IP65)

Cell constant K

0.01 (cm

-1

)

0.1 (cm

-1

)

1.0 (cm

-1

)

0.01 (cm

-1

)

Accuracy

< 2%

< 2%

< 2%

< 2%

Measurement range

0.01—200 μS.cm

-1

0.1 μS—2 mS.cm

-1

1 μS—20 mS.cm

-1

0.01—200 μS.cm

-1

Temperature
response

< 30 seconds

< 30 seconds

< 30 seconds

< 30 seconds

Maximum
temperature

150 °C

150 °C

150 °C

150 °C

English 3