Hach-Lange POLYMETRON 9500 Conductivity Module User Manual User Manual

Page 21

Advertising
background image

Table 4 Warning list for conductivity sensors (continued)

Warning

Description

Resolution

TEMP TOO HIGH

The measured temperature is >
200 °C

Make sure that the sensor is
configured for the correct
temperature element.

TEMP TOO LOW

The measured temperature is <
-20 °C

CAL OVERDUE

The Cal Reminder time has expired

Calibrate the sensor.

NOT CALIBRATED

The sensor has not been calibrated

Calibrate the sensor.

REPLACE SENSOR

The sensor has been in operation >
365 days

Calibrate the sensor with a
reference solution and reset the
sensor days. Refer to

Sensor

diagnostic and test menu

on page 19. If the calibration fails,
call technical support.

CAL IN PROGRESS

A calibration was started but not
completed

Return to calibration.

OUTPUTS ON HOLD

During calibration, the outputs were
set to hold for a selected time.

The outputs will become active after
the selected time period.

WRONG LINEAR TC

The user-defined linear temperature
compensation is out of range

The value must be between 0 and
4%/°C; 0 to 200 °C.

WRONG TC TABLE

The user-defined temperature
compensation table is out of range

The temperature is above or below
the temperature range defined by
the table.

WRNG USER CONC TABLE

The concentration measurement is
outside of the range of the user
table

Make sure that the user table is set
for the correct measurement range.

WRNG BLT-IN TEMP TABLE

The measured temperature is
outside of the range of the built-in
temperature compensation table

Make sure that the temperature
compensation is configured
correctly.

WRNG BLT-IN CONC TABLE

The concentration measurement is
outside of the range of the built-in
concentration table

Make sure that the concentration
measurement is configured for the
correct chemical and range.

English 21

Advertising