2 hardware description, 3 model identification system, Specifications – Hach-Lange ORBISPHERE 410 User Manual User Manual

Page 16

Advertising
background image

14

Specifications

2.2 Hardware description

The instrument hardware is made of one main board, and one measurement board for the
measurement channel (= the sensor).

The main board includes the controls for power, display, the touch screen, the barometric
sensor, the alarms, and communication ports. The measurement board performs measurements
and executes commands from the main board. It holds the "Analog output" and "Relays" that
send information to external systems.

A hardware watchdog is activated at program start up, to check that the system is not frozen
(i.e. infinite loop, system crash, etc.). If the watchdog is not refreshed by the software every
minute, the measurement display, the relays and the analog output are frozen up to 2 minutes.
Then the reset shuts down the instrument for 10 seconds and the start-up procedure is
performed. At the same time all the hardware (sensor, measurement board) are reset.

2.3 Model identification system

The analyzer identification number and the instrument serial number are located on the label on
the back panel, and can be found on order confirmation and invoice papers.

COMMUNICATION

Options

RS-485 or PROFIBUS-DP (optional)
USB client
USB host
Ethernet 10/100 Base-T

GENERAL

Thermal cut off

Prevents ageing of sensors when exposed to high temperatures

SIZE AND WEIGHT

Wall and pipe mount
(H x D x W)

236.5 x 160 x 250 mm - weight 3.8 kg
9.31 x 6.30 x 9.84 ins. - weight 8.38 lbs

Panel mount: Face (housing)
(H x D x W)

156 (123) x 250 x 220 (214) mm - weight 2.9 kg
6.14 (4.84) x 9.84 x 8.86 (8.43) ins. - weight 6.39 lbs

A Oxygen EC sensor

C Ozone EC sensor

D Carbon dioxide TC sensor

W Wall

P Panel

1

100-240 VAC

2

10-30 VDC

C 0/4-20 mA

V 0-5 V

0

RS-485

1

PROFIBUS-DP / RS-485

0

Standard

K Korean language

4

1

0

/

/

0

0

0

Advertising