Maintenance and Troubleshooting

14.3 List of events and alarms

14.4 Storage, handling and transportation

Protect the instrument against the elements: rain, splashing, direct sunlight, etc.

A properly packaged instrument can be stored and transported at a temperature -20°C to +70°C
and relative humidity up to 80%. Best practice for packing the instrument for transportation is to
reuse the original packaging in which the instrument was first delivered. The instrument should
be stored in suitable premises, free of dust, condensation and chemical evaporation.

In cold weather, avoid sudden temperature change (like when entering a warm room) and give
the instrument enough time to adapt to the ambient temperature in order to avoid condensation
inside.

To clean the instrument, wipe the housing clean with a cotton cloth or tissue. Always clean the
instrument before storage. Pay attention not to scratch the surface of the display to retain good
clarity over time.

Table 4 List of Events

Event

type

Name

Description

Bit mask

value (32

bits long)

In

for

m

atio

n

Measure

Normal measurement mode.

0x00000000

Filter enabled

The gas measurements are filtered.

0x00000001

Sample measurement

The sample measurement is started.

0x00000002

Meas. not ready

The measurement is not ready (e.g. at startup)

0x80000000

Autotest in progress

The autotest is running.

0x00100000

Autotest failed

The autotest has failed.

0x00200000

Alarm snooze

The alarm snooze is ON.

0x00000004

W

a

rn

ing

Calibration

Channel in calibration.

0x00000008

Hold

The measurement is frozen.

0x00400000

Alarm low low

The gas concentration below the Alarm LowLow limit.

0x00000010

Alarm low

The gas concentration is below the Alarm Low limit.

0x00000020

Alarm high

The gas concentration is above the Alarm High limit.

0x00000040

Alarm high high

The gas concentration is above Alarm HighHigh limit.

0x00000080

Calibration required

The calibration of the sensor is required.

0x00000100

Service required

The sensor requires a service.

0x00000200

TC hold

TC measurement on hold

0x08000000

Continuous purge on

Continuous purge is on

0x10000000

Al

ar

m

Channel disabled

The channel has been disabled.

0x00000400

Channel out

The measurement board has been disconnected (or does not answer).

0x00000800

Sensor out

The sensor has been disconnected.

0x00001000

PROFIBUS-DP

The PROFIBUS-DP module has not received measurements from the
instrument for 30 seconds.

0x00080000

Thermal cut off

The temperature is above the thermal cut off.

0x00004000

Ramp limit violated

V3 voltage outside range -3V to +8V

0x04000000

NOTICE

Never use liquids such as oil, benzene, or detergents for cleaning the instrument. A mild glass
cleaner can be used to remove greasy stains.