4 http/tcp-ip, 1 overview, 2 pc interface – Hach-Lange ORBISPHERE 410 User Manual User Manual

Page 77: 1 overview 9.4.2 pc interface

Advertising
background image

75

Communication Menu

9.4 HTTP/TCP-IP

9.4.1 Overview

When activated this option downloads data from the instrument directly to a web page that can
be accessed from a PC. To be able to use this option, the instrument must be connected to the
network (specifically Connector P3 - see details in

Main board connections on page 29

) and

the network must have a DHCP server installed.

Check the Enable server box to enable the web server communication link.

Enter the Server name for the instrument. This is free format text and should typically be
used to identify the instrument.

Check the Enable security box if you require a password to be entered on the PC to access
the web page.

If any of the details on the previous screen have been changed, a warning message will be
displayed as illustrated left.

The changes must be confirmed, after which the instrument must be powered down and
powered up again for the changes to take effect.

Note:

The IP Diagnostics button at the bottom of the screen is for use by experienced IT personnel only to

help resolve any communications problems.

9.4.2 PC interface

Once the server has been enabled and the interface information set up, access the information
by launching an internet browser and typing “http://” followed by the server name that has been
assigned to the instrument, in the address box as illustrated below:

If the enable security option has been checked on the
instrument, you will be required to enter a username
and password on your PC to gain access to the web
page.

The username and password must be a valid
username/password combination that has been set up
on the instrument (see also

User management on

page 80

on how to set up users on the instrument).

Domain information is not required.

Advertising