4 measurement board, Installation 3.7.4 measurement board – Hach-Lange ORBISPHERE 410 User Manual User Manual

Page 32

Advertising
background image

30

Installation

3.7.4 Measurement board

The different measurement boards for the EC and TC sensors are illustrated in

Figure 14

and

Figure 15

below. The type of board is easily identified by the color of the J8 connector. For EC

boards this connector is colored orange, and for TC boards it is colored black.

The colors indicated are the wire colors in the sensor cable.

NOTICE

It is extremely important that sensors are connected to the correct measurement board.
Connecting a TC sensor to an EC measurement board (and vice versa) will cause irreparable
damage to the measurement board.

Figure 14 EC Measurement board

Figure 15 TC Measurement board

Figure 16 Connector J7

Figure 17 Connector J8

Connector J7 (inputs & outputs)

Measurement alarms relays

1.

Common

2.

Output relay 1

3.

Output relay 2

4.

Output relay 3

Note:

Relays N.O. or N.C. depends on the jumper

position on the relays (see

Measurement alarm

relays on page 31

).

Analog current (or voltage) outputs

5.

GND

6.

Output 1

7.

Output 2

8.

Output 3

Digital inputs

9.

EC sensor: Not used
TC sensor: Hold input (short circuit to pin
12)

10.

Not used

11.

Not used

Advertising