Installation

Note:

To change the type of sensor (e.g. from a 31xxx sensor to a 31xxxS smart sensor) contact your local

Hach Lange representative.

3.8 Measurement alarm relays

Connector J8 (sensor)

Note:

Remember, this connector is colored orange for EC sensors and black for TC sensors.

A1100 EC sensor

31xxx EC sensor

31xxxS smart EC

sensor

TC sensor

Cable

sensor

1)
2)
3)
4)
5)
6)
7)
8)
9)

10)

Guard electrode
RS485A+
Thermistor A
Anode electrode
RS485B-
Thermistor B
GND
+ 5V
Cathode electrode
Not used

Guard electrode
Not used
Thermistor A
Anode electrode
Not used
Thermistor B
Not used
Not used
Cathode electrode
Not used

Guard electrode
I2C-SCL
Thermistor A
Anode electrode
I2C-SDA
Thermistor B
GND
+ 5V
Cathode electrode
Not used

GND for power
V2 signal
Solenoid
Relay coil
+12V power
+24V power
V3 signal
GND for signal
-5V power
Temperature

Yellow

Pink

Grey

Red

Purple

White

Black

Green

Blue

Brown

Figure 18 Output Relays

The three output relays are located on the measurement
board (see also

Figure 14

and

Figure 15 on page 30

).

They can be individually configured to Normally Open (NO) or
to Normally Closed (NC) by physically moving the jumper on
each relay. The picture example on the left is for the EC
measurement board (the positions are different for the TC
measurement board):

Upper relay is set to NC

Middle relay is set to NO

Lower relay is shown with no jumper, to show the 3 pins.

Note:

For all measurement boards, J4 is relay 1, J5 is relay 2 and J6
is relay 3