Installation

3.4.3 Connection panel (bottom of instrument)

Alternative instrument mounting procedure
When it is not convenient to work from the back of the panel, the instrument can be connected
before fitting in the panel.

1.

Install the panel support frame in the panel opening

2.

Slip the cables through the panel opening

3.

Remove the instrument cover

4.

Slip the cables through the instrument cover

5.

Slip the cables through the instrument back panel cable glands

6.

Connect the cables to the instrument electronic boards

7.

Tighten the cable glands

8.

Reinstall the instrument cover

9.

Install the instrument in the panel opening

1.

Power cable. The type of connection will
vary depending on the instrument
specification (see

Connection to mains

power supply on page 26

)

2.

Ethernet cable gland

3.

USB-A host connector for mass storage
device

4.

USB-B client 4 pin connector. Use the
adapter cable (see

USB-B client adapter

cable on page 26

)

5.

Input/Output 1 cable gland

6.

Sensor cable gland

7.

Input/Output 2 cable gland

Figure 8 Panel mount connection panel