Services Menu

13.2 Language selection

13.3 Clock

13.4 Screen

13.5 Buzzer

13.6 Boards info

Amplifiers (TC sensor
only)

This feature will display the measured values for V2 and V3 along with the
recommended values. It is advisable to put the sensor in continuous purge
mode during this operation by pressing the Continuous purge button.
If the measured and recommended values differ by a large margin, please
contact your Hach Lange representative for advice.

Language selection

Check the language as required and restart the instrument to apply the change. The instrument will
restart in the language selected.

Clock

Type in each appropriate box the actual time and date, and select the display format for them.

Screen

Screen calibration

This screen allows you to adjust the click position corresponding to the
displayed buttons. Use it if ever the sensitive areas are no longer properly
aligned with the buttons on display.
Place the stylus right on the cross when asked and proceed. User will be
asked to press on the screen to accept the new setting. If not, the new
setting is not recorded and no change is made.

Screen brightness

Press the up or down arrow to increase or decrease the screen brightness.
Four levels are pre-set: Max, Comfortable, Standard (default setting) and
Economy

. Press OK when finished.

Note:

This can also be called through the brightness icon on the

main display.

Buzzer

Adjust the sounds available on the instrument.
When “screen tap” is enabled, a click sound is heard each time the screen is touched.
The instrument alarm sound can be enabled or disabled to suit the application. The sound type can also
be adjusted. Press the Test button to test the adjustment made. Press again to stop.

Boards info

Main board info

For reference, this display gives information on the instrument model,
software version and instrument ID.

Measurement board info

For reference, this display gives information on the sensor measurement
board hardware and software.
The Model field indicates if the sensor is an EC or TC sensor.
Pressing the Analog outputs corr. button displays the correction factor that
is applied to the analog outputs.
Pressing the Amplifiers corr. button displays the value of the actual
correction factor on the amplifiers.

Sensor parameters

For reference this display gives information on the sensor model and type,
last calibration, settings and behavior.