Expanded programs

7. Inset sample cell 1 (

Figure 3, page 19

) in cell compartment

(1) a press Next.

8. Follow the dialog indicated.

9. After the fifth test cell, the photometrical accuracy is

displayed in a log.

6.8.3.3

Wavelength check

The Wavelength Check test is used to check wavelength
accuracy at 807 nm.

1. Press Wavelength Check in the "Optical Checks" menu.

2. Insert the cell adapter (A) for 10-mm rectangular cells in the

cell compartment (2).

3. Place the analysis cell (Neodym or BG20/2) in the cell

adapter.

4. Close the cell compartment cover.

5. Press Start.

The result will be displayed.

6. Compare the result with the analysis cell required values.

7. Press Cancel to return to Optical Checks.

6.8.3.4

Stray light check

The Stray Light test is used to measure the stray light in the
instrument at 340 nm.