Expanded programs – Hach-Lange DR 3900 User Manual User Manual

Page 85

Advertising
background image

85

Expanded programs

Content and definition of the keys

[Z] key / Zeroing

1. Press [Z] to program a zero reading.

2. Press OK to confirm.

3. Press New.

4. Press Zeroing and enter the wavelength for which the zero

reading is to be performed.

5. Press OK and confirm the entry.

6. Press OK.

7. If zero readings are to be carried out at a number of

wavelengths, repeat the steps starting with 1 for each
wavelength.

Note: The entered reading sequence is displayed.

Process Timer key

1. Press New.

2. Press Process timer to enter wait, reaction or handling

times.

3. Press OK and confirm the entry.

4. Press OK.

Note: This time is integrated into the measurement process.

Note: The entered reading sequence is displayed.

[R] key / Reading...

1. Press New.

2. Press [R] to program a reading of the substance that is to

be analyzed.

3. Press OK to confirm.

4. Press New.

5. Press Read...and enter the wavelength for which the

measurement is to take place.

6. Press OK and confirm the entry.

7. Press OK.

8. If readings are to be completed at a number of wavelengths,

repeat the steps starting with 1 for each wavelength.

Note: The entered reading sequence is displayed.

Advertising