Troubleshooting

Under measuring range

The measured absorbance is below the
calibration range of the test

If possible, select a test with a lower
reading range or use a cuvette with a
longer path length

Possible interference by:

Interference Check

The analysis is possibly erroneous due
to interferences.

Possible interference from:

Interference Check

The analysis is possibly erroneous due
to interferences.

Next service is due!

The service reminder is activated.

Contact technical support for an
inspection of the instrument. For setting
the service reminder, refer to

section 6.8.7, page 123

Negative result!

The calculated result is negative

Check concentration of sample

Network switched off.

Network setup off, when accessing the
Instruments homepage via the sidebar

Activate the online connection.

Remote server cannot be reached.

Error during network setup

Make sure that the instrument is
connected to the network.

Unstable lighting conditions!

Avoid direct sunlight at the measuring
location.

Insufficient memory for update .

Error during update.

Select a memory with more space.

System check incorrect!

Measurement of air values failed

Switch the instrument off and then back
on again. If the system check is not
successful, contact technical support.

Temperature too high.
Measurement not possible!

Switch of the instrument and allow it to
cool for a few minutes. If necessary,
move it to a cooler place.

Update file is faulty.

Error during update.

Save the update file again and repeat
the procedure.

USB memory stick is not connected.

Update not possible.

Check the USB memory stick.

Web server cannot be reached.

Instruments homepage cannot be
reached

Try the connection again later.
Make sure that the instrument is
connected to the network.

Error displayed

Definition

Resolution