3 exchange of the replaceable cell compartment – Hach-Lange DR 3900 User Manual User Manual

Page 130

Advertising
background image

130

Maintenance

4. Install the lamp holder in the instrument.

Note: Take care that the lamp cable is not pinched by the cable
guide.

5. Using a cross-head screwdriver, tighten the two lamp holder

screws.

6. Replace the lamp cover.

7. Using a slot-head screwdriver tighten the cover screws.

8. Connect the power cable to the instrument.

9. Reset the lamp operating time

section 6.8.8, page 123

.

7.3

Exchange of the replaceable cell compartment

If there is severe soiling, the complete cell compartment (2) can
be replaced.

1. Switch off instrument.

2. Remove the power cable from the instrument.

3. Using a cross-head screwdriver, undo the two screws on the

floor of the cell compartment.