Section 1 specifications – Hach-Lange DR 3900 User Manual User Manual

Page 9

Advertising
background image

9

Section 1

Specifications

These are subject to change without notice!

Performance specifications

Display mode

Transmittance (%), Absorbance and Concentration

Source lamp

Halogen lamp

Wavelength range

320–1100 nm

Wavelength Accuracy

± 1.5nm (wavelength range 340–900 nm)

Wavelength reproducibility

± 0.1 nm

Wavelength resolution

1 nm

Wavelength calibration

Automatic

Wavelength selection

Automatic, based on method selection

Scanning speed

≥ 8 nm/s (in steps of 1 nm)

Spectral bandwidth

5 nm

Photometric measuring range

± 3.0 Abs (wavelength range 340–900 nm)

Photometric accuracy

5 m abs at 0.0–0.5 abs
1 % at 0.50–2.0 Ext

Photometric linearity

< 0.5 % to 2 Abs
≤1 % at > 2 Abs with neutral glass at 546 nm

Stray light

< 0.1 %T at 340 nm with NaNO

2

Data log

2000 readings (Result, Date, Time, Sample ID, User ID)

User programs

100

Physical and environmental specifications

Width

350 mm (13.78 in)

Height

151 mm (5.94 in)

Depth

255 mm (10.04 in)

Mass

4200 g (9.26 lb)

Ambient operating requirements

10–40 °C (50–104 °F), maximum 80 % relative humidity (without condensate
formation)

Ambient storage requirements

–40–60 °C (–40–140 °F), maximum 80 % relative humidity (without condensate
formation)

Additional technical data

Power connector via external power
supply

Input: 100–240 V/50–60 Hz
Output: 15 V/40 VA

Interfaces

Use only shielded cable with maximum length of 3 m:
2× USB type A
1× USB type B
Use only shielded cable (for example STP, FTP, S/FTP) with maximum length of
20 m:
1× Ethernet

Housing rating

IP40 (excluding interfaces and power supply)

Protection class

Class I

Advertising