Standard programs

Enter the measurement process

For detailed information on entering the measurement process,
refer to

section 6.1.2, page 82

.

Check the process specification in advance to determine whether
this point needs to be changed.

1. Highlight the Measurement process line in the data

overview and press Edit.

2. Press Edit, select the sequence to be edited and press

Delete.

3. Press New and use the alphanumeric keypad to enter the

process indicated in the process specification.

Enter the Formula, Concentration Units, Designation,
Measuring Ranges

For detailed information on entering the Formula, refer to

section 6.1.2, page 82

.

Check the process specification in advance to determine whether
this information needs to be changed.

1. Highlight the Formula line in the data overview and press

Edit.

2. Select the formula; press Edit, and enter the data (refer to

process specification) (for C1=, C2=, units, name,
measurement range limits, etc.). Press OK to confirm the
entry.

Enter the Variables (Factors, Wavelength and Coversion
Factors)

For detailed information on entering the Variables, refer to

section 6.1.2, page 82

.

Check the process specification in advance to determine whether
this point needs to be changed.

1. Highlight the Variables line in the data overview and press

Edit.

2. Select the variable, tap on Edit, and enter the data (refer to

process specification) with the alphanumeric keypad (for F1,
F2,

λ 1, U1 etc.). Confirm each entry with OK.

Abbreviation of Variables:

F1: Factor 1
F2: Factor 2
λ1: Wavelength 1

U1: Conversion factor 1 for the first chemical form
U2: Conversion factor 2 for additional chemical forms, etc.