Start Up

4.5

Setting the display

It is possible to custom-define the first two rows of the display. In the standard setting, the
first line shows the main measured variable and the second line shows the date/time.

1. Select

MENU

, confirm with

ENTER/ON

.

2. Select

DISPLAY

, confirm with

ENTER/ON

.

3. Select

ROW 1

, confirm with

ENTER/ON

.

4. Select option for

ROW 1

, confirm with

ENTER/ON

.

Measurement value

Homogeneous

Battery

Time

5. Select

ROW 2

, confirm with

ENTER/ON

.

6. Select option for

ROW 2

, confirm with

ENTER/ON

.

7. Press

CLEAR/OFF

to return to the

MENU

.

Note: Press CLEAR/OFF again to return to the main menu.

4.6

Setting the units

The following measurement units are available for turbidity and solids:

Turbidity units (TU): FNU, NTU, EBC

Solids units (DS1 to DS4): ppm, mg/L, g/L, %

1. Select

MENU

, confirm with

ENTER/ON

.

2. Select

UNITS

, confirm with

ENTER/ON

.

3. Select the desired calibration curve (C-TU, C-DS1, C-DS2, C-DS3 or C-DS4) and

confirm with

ENTER/ON

.

4. Select the desired measurement unit and confirm with

ENTER/ON

.

5. Press

CLEAR/OFF

to return to the

MENU

.

Note: Press CLEAR/OFF again to return to the main menu.

4.6.1

Sludge blanket level

To determine the sludge blanket level, set the units to mg/L, g/L or %. Select an integration
time of 0 seconds (refer to

section 4.9, page 21

) to increase the response time. Calibrate

the measurement system and slowly lower the sensor into the clarifier. Once the sludge
blanket level is reached, the solids concentration will increase significantly. The distance
of the sludge blanket from the surface of the water can then be read off from the meter
markings on the probe cable.