1 selecting the calibration curve, 2 start the measurement, 3 interrupt and resume the measurement – Hach-Lange TSS PORTABLE Handheld Instrument for Turbidity_Suspended Solids Manual User Manual

Page 27: 4 stop measurement, 3 show the data in the display, 4 delete data for a calibration curve, Operation 5.2.1 selecting the calibration curve

Advertising
background image

27

Operation

5.2.1

Selecting the calibration curve

Before starting the measurement, the calibration curve that corresponds to the
measurement point must be selected.

1. Select

MENU

, confirm with

ENTER/ON

.

2. Select

CALIBRATE

, confirm with

ENTER/ON

.

3. Selectthe curve shown and confirm with

ENTER/ON

.

4. Select the desired curve with the navigation key

UP/DOWN

and confirm with

ENTER/ON

.

5. Select

CLEAR/OFF

twice to return to the main menu.

5.2.2

Start the measurement

When starting the measurement, the measurement is carried out continuously and the
current measurement value is saved at intervals.

1. Select

MEASURE

, confirm with

ENTER/ON

.

2. Select

START

and confirm with

ENTER/ON

.

5.2.3

Interrupt and resume the measurement

1. Press

CLEAR/OFF

for 2 seconds.

The instrument is switched off and the measurement is interrupted.

2. Press

ENTER/ON

for 2 seconds.

The instrument switches on and the measurement continues with the same settings
as before the instrument was switched off.

5.2.4

Stop measurement

1. Select

MEASURE

, confirm with

ENTER/ON

.

2. Select

STOP

, confirm with

ENTER/ON

.

5.3

Show the data in the display

The display shows the stored data for the corresponding calibration curves.

1. Select the desired calibration curve (refer to

section 5.2.1, page 27

).

2. Select

DATA

, confirm with

ENTER/ON

.

3. Select

DISPLAY DATA

and confirm with

ENTER/ON

.

Only the stored data for the selected calibration curve is displayed.

5.4

Delete data for a calibration curve

1. Select the desired calibration curve (refer to

section 5.2.1, page 27

).

2. Select

DATA

, confirm with

ENTER/ON

.

3. Select

DELETE MEP

and confirm with

ENTER/ON

.

Only the stored data for the selected calibration curve is deleted.

4. Select

YES

, confirm with

ENTER/ON

.

Note: NO aborts the erasing procedure.