7 calibration, 8 defining the measurement points, 1 overview of calibration curves – Hach-Lange TSS PORTABLE Handheld Instrument for Turbidity_Suspended Solids Manual User Manual

Page 19: 2 measurements, 1 practical example for solids measurement, 7 calibration 4.8 defining the measurement points

Advertising
background image

19

Start Up

4.7

Calibration

For turbidity measurement, a standard curve C-TU is already stored in the probe.
Calibration is not required.

However, for solids measurement, calibration is required in order to set the rough
measurement signals to a calibrated display (refer to

section 5.1, page 23

). A precise

solids measurement is not possible without calibration. There are four calibration curves
available: C-DS1, C-DS2, C-DS3, C-DS4. These curves can be allocated to the individual
measurement points.

4.8

Defining the measurement points

4.8.1

Overview of calibration curves

C-TU = Turbidity curve (standard curve)

C-DS1 = solids curve 1 (calibration curve)

C-DS2 = solids curve 2 (calibration curve)

C-DS3 = solids curve 3 (calibration curve)

C-DS4 = solids curve 4 (calibration curve)

4.8.2

Measurements

4.8.2.1

Practical example for solids measurement

Note: The type and composition of the solid particles of a medium can differ greatly. For this reason,
it is not possible to define a general standard, and different measurement media must be calibrated
specifically in each case.

For the solids measurement, calibration is required. There are four curves available:
C-DS1, C-DS2, C-DS3, C-DS4. Each curve can be assigned to any measurement point.
In the example, curve C-DS1 is selected.

1. Connect the probe to the instrument.

2. To switch on the instrument, press

ENTER/ON

for 2 seconds.

Select calibration curve

3. Select

MENU

, confirm with

ENTER/ON

.

4. Select

CALIBRATE

, confirm with

ENTER/ON

.

5. Select

CURVE C-TU

and confirm with

ENTER/ON

.

Note: Curve C-TU is the default setting.

6. Select calibration curve

C-DS1

with the navigation keys

UP/DOWN

and confirm

with

ENTER/ON

.

The instrument reads the probe data and info text appears on the display.

Advertising