3 manual calibration value correction, 4 delete calibration point, 5 reset to default calibration – Hach-Lange TSS PORTABLE Handheld Instrument for Turbidity_Suspended Solids Manual User Manual

Page 26: 2 measure

Advertising
background image

26

Operation

5.1.3

Manual calibration value correction

1. Select

MENU

, confirm with

ENTER/ON

.

2. Select

CALIBRATE

, confirm with

ENTER/ON

.

The instrument reads the probe data.

3. Select the curve shown and confirm with

ENTER/ON

.

4. Select the desired curve and confirm with

ENTER/ON

.

5. Select the desired point and confirm with

ENTER/ON

.

6. Overwrite the existing calibration point.

7. Enter the value for the calibration point using the navigation keys and confirm with

ENTER/ON

.

Navigation key

LEFT/RIGHT

: Jump to next/previous decimal place

Navigation key

UP/DOWN

: Change number

5.1.4

Delete calibration point

Same procedure as in

section 5.1.3

. Set the concentration of the calibration point to the

value 0.000 and confirm with

ENTER/ON

. The calibration point is deleted.

5.1.5

Reset to default calibration

If the turbidity calibration was changed, it can be reset to the default calibration in
accordance with ISO 7027.

Note: Stop the current measurement before the default calibration is reset.

1. Select

MENU

, confirm with

ENTER/ON

.

2. Select

CALIBRATE

, confirm with

ENTER/ON

.

The instrument reads the probe data.

3. Select turbidity curve C-Tu and confirm with

ENTER/ON

.

Note: If turbidity curve C-Tu is not shown in the menu, select the curve shown and confirm with
ENTER/ON. The selection menu for the curves is shown. Select curve C-Tu and confirm with
ENTER/ON. Curve C-TU is activated.

Open the selection menu for curve C-TU with ENTER/ON, select RESET and confirm with
ENTER/ON.

4. Press navigation key

UP/DOWN

, select

RESET

and confirm with

ENTER/ON

.

The DS curves can also be reset to the default calibration (1-point calibration at 4 g/L).
However, a precise determination of the solids concentration can only be obtained through
calibration (refer to

section 5.1, page 23

).

5.2

Measure

Note: In order to obtain precise solids measurement results, you must calibrate the instrument
before the first measurement is taken.

Without any calibration of the DS curves, the instrument returns to 1-point default calibration at
4 g/L. Deviations from the actual measurement value may occur.