Start Up

4.8.2.2

Practical example for turbidity measurement

A standard curve C-TU is stored for the turbidity measurement.

1. Connect probe to instrument.

2. To switch the instrument on, press

ENTER/ON

for 2 seconds.

3. Select

READ

and confirm with

ENTER/ON

.

4. Place probe in measuring medium.

5. Select

START

and confirm with

ENTER/ON

.

Note: The data is automatically saved every minute, max. 290 measurement points.

6. To stop the measurement, select

READ

and confirm with

ENTER/ON

.

7. Select

STOP

, confirm with

ENTER/ON

.

Delete the saved measurement points:

8. Select

DATA

and confirm with

ENTER/ON

.

9. Select

DELETE MEP

and confirm with

ENTER/ON

.

10. Select

YES

and confirm with

ENTER/ON

.

4.9

Setting the integration time

Note: The measurement values are averaged out over the integration time, which results in a
smooth measurement signal. For the sludge blanket level measurement (refer to

section 4.6.1,

page 18

), select the integration time 0 seconds.

Upon delivery, the default value is set to 30 seconds. The integration time can be set to
between 0 and 1000 seconds.

1. Select

MENU

and confirm with

ENTER/ON

.

2. Select

EXTRAS

and confirm with

ENTER/ON

.

3. Select

INTEGRATION

and confirm with

ENTER/ON

.

4. Change the displayed value with the navigation keys.

5. Confirm each entry with

ENTER/ON

.

6. Press

CLEAR/OFF

to return to the

MENU

.

Note: Press CLEAR/OFF again to return to the main menu.