20

Start Up

Calibrate curve C-DS1:

7. Select

MEMORY

and confirm with

ENTER/ON

.

8. Lower the probe into a container that contains a homogeneous sample.

9. Select

POINT 1

and confirm with

ENTER/ON

while stirring the measuring medium with

the probe.

The distance between the probe head and the walls and base of the container must
always be more than 70 mm (2.76 in.).

It takes 5 to 20 seconds to record the calibration point; then the selection menu for
calibration appears. While recording, the display shows the note "Memory".

10. Take the probe out of the container and clean it.

11. Determine the solids content of the sample in the laboratory.

12. Select

*POINT 1

and confirm with

ENTER/ON

.

13. Enter the laboratory value with the navigation keys and confirm with

ENTER/ON

.

Navigation key

LEFT/RIGHT

: Jump to next/previous decimal place

Navigation key

UP/DOWN

: Change number

The laboratory value is saved. The main menu is shown.

Measure:

14. Select

MEASURE

, confirm with

ENTER/ON

.

15. Select

START

and confirm with

ENTER/ON

.

Note: Data is automatically saved every minute, max. 290 measurement points.

Stop measurement:

16. Select

MEASURE

, confirm with

ENTER/ON

.

17. Select

STOP

, confirm with

ENTER/ON

.

Delete the saved measurement points:

18. Select

DATA

, confirm with

ENTER/ON

.

19. Select

DELETE MEP

and confirm with

ENTER/ON

.

20. Select

YES

and confirm with

ENTER/ON

.