Flow input – Liquid Controls SP4000 User Manual

Page 65

Advertising
background image

62

SP4000 Flow Computer

The DP-Factor describes the relationship between the flowrate and the
measured differential pressure. The flowrate is computed according to one of
the three following equations, depending on the selected flow equation:

Steam (or gas) mass flow:

Liquid volume flow:

Gas corrected volume flow:

M

Mass flow

Q

Volumetric flow

Q

REF

Corrected volumetric flow

K

DP

DP-Factor

ε

1

Gas expansion factor (Y

1

)

T

Operating temperature

T

CAL

Calibration temperature

∆

p

Differential pressure

ρ

Density at flowing conditions

K

ME

Meter expansion coefficient x 10

–6

ρ

REF

Reference density

DP FACTOR

6.8
FLOW INPUT
(Continued)

FLOW INPUT

2 •

∆

p •

ρ

1 – K

ME

• (T – T

CAL

)

√

K

DP

ε

1

•

•

M =

(2 •

∆

p)

/ρ

(1 – K

ME

• (T – T

CAL

))

√

K

DP

•

Q =

2 •

∆

p •

ρ

(1 – K

ME

• (T – T

CAL

))

√

Q

REF

=

K

DP

ε

1

•

•

ρ

REF

•