Luminex 100 IS Version 2.2 User Manual

Page 124

Advertising
background image

Luminex 100 IS User Manual Version 2.2

x

MAP Technology

6 - 70

PN 89-00002-00-069 Rev. A

Change Start

Acquiring Data

Location in Multi-

Batches

You can change the location for multi-batches to begin acquiring
data.

‹

To change the locations for beginning acquisition for batches in
a multi-batch:

1. Click Create New Batch or Create New Multi-Batch on the

toolbar. The Luminex Multi-Batch Setup dialog box appears.
Refer to “Create a New Batch” on page 6-55 or “Create a Multi-
Batch” on page 6-61 for more information.

2. From the microtiter plate representation of the Luminex Batch

Setup or Luminex Multi-Batch Setup dialog box, click the first
well for sample acquisition. See Figure 66.

1. First batch

2. Separator Line

3. Second batch

Figure 66. Selecting Multi-Batch Starting Data

3. Drag the highlighted well to the location on the microtiter plate

where you want to begin acquiring data. For a multi-batch ensure
each additional batch is in its desired location. The display
updates to show the wells for acquisition in its new location. The