Sample error messages – Luminex 100 IS Version 2.2 User Manual

Page 197

Advertising
background image

x

MAP Technology

Troubleshooting

PN 89-00002-00-069 Rev. A

8 - 7

If you power the off, you must calibrate and verify the system
when the system properly connects.

Locked Out

•

Close all software applications other than the Luminex 100 IS
2.2 software.

Refill Sheath

A warning displays in red under the temperature and pressure gauges
on the Run Batch tab.

•

Refill the sheath container and select Resume.

Running: Sheath

Empty

•

Replace or fill the sheath fluid container with sheath fluid. From
the Maintenance tab, click Prime 2 times.

Sample Error Messages

The following error messages are displayed:

•

On the Diagnostics tab in the Message Log

•

In Reports

•

On the Errors tab in a batch analysis.

Insufficient Bead

Count

•

Do not select Cancel during sample acquisition. This causes an
error message to occur.

•

Verify that you select the correct template.

•

If applicable, verify that all of the tests (beads) were added to the
samples.

•

Verify that the samples were properly mixed before they were
dispensed into the wells.

•

Verify that you loaded the sample to the correct well on the
plate.

•

Verify that the correct sample volume is added to the well.

•

Verify that the samples were completely resuspended prior to
running on the analyzer.

•

Remove and sonicate or flush the Luminex XYP instrument
Sample Probe to remove possible clogs.

•

Verify that the Luminex XYP instrument Sample Probe is
correctly aligned.

•

Complete the following analyzer functions from the Maintenance
Tab: Backflush, Drain, Alcohol Flush and Water Wash.

•

Run the system xMAP control microspheres.

Note:

Do not install

unapproved software
applications on the Luminex
100 IS 2.2 system.