Luminex 100 analyzer problems – Luminex 100 IS Version 2.2 User Manual

Page 200

Advertising
background image

Luminex 100 IS User Manual Version 2.2

x

MAP Technology

8 - 10

PN 89-00002-00-069 Rev. A

•

Verify that you selected the correct template.

•

Verify that lot information is entered correctly.

•

Verify that the system is functioning properly by running system
controls.

•

If the system controls pass, contact the kit manufacturer.

Luminex 100 Analyzer
Problems

Power On

Indicator Light Is

Not On

•

Verify that the Luminex 100 analyzer is powered on.

•

Verify that the system is plugged into an outlet with current.

•

If using a surge protector, verify that the power strip is On.

•

Remove and reconnect the cable.

•

Verify the fuses (see “Fuses” on page 7-11).

PC Is Not Powered

On

•

Verify that the PC voltage is set correctly by checking the back
of the PC. The options are 115 volts and 230 volts.

•

Verify that the system is plugged into an outlet with current.

•

If using a surge protector, verify that the power strip is On.

•

Remove the cable and reconnect.

•

Power on the PC.

Sheath Fluid Is

Leaking Onto the

Plate Holder

•

Remove and sonicate or flush the Luminex XYP instrument
Sample Probe. Load a new plate onto the plate holder.

•

Complete the following analyzer functions from the Maintenance
tab: Backflush, Drain, Alcohol Flush and Water Wash.

•

Empty the waste container. Prime the system and check if sheath
fluid is dripping into the waste container. Contact Luminex
Customer Support with this information.

•

If using the filter bottom plates, adjust the Sample Probe height
up to prevent the probe from puncturing the filter.

Progress Bar

Indicates No

Events Being

Collected During

Calibration or

Verification

•

Verify that the Luminex 100 analyzer and the Luminex XYP
instrument are connected properly.

•

Empty the waste container if it is completely filled.

•

Verify that the tubing from the sheath fluid and waste containers
are connected properly.

•

Verify that you properly mixed the system calibrators or controls
before dispensing into the wells.

•

Verify that you added the system calibrators or controls to the
correct well on the plate.

•

Verify that the volume added to the well is correct.