Calibrator xMAP microspheres are used to normalize the settings for
the reporter channel, both classification channels, and the doublet
discriminator channel for the Luminex 100 IS 2.2 system.

The control xMAP microspheres are used to verify the calibration
and optical integrity for the system.

You must calibrate at least once a month and:

following installation

if the system is powered off

if the system is moved

if a part is replaced

if the delta calibration temperature shown on the system monitor
is more than ±3 degrees.

Each step usually takes less than one minute. You must run xMAP
controls after each calibration.

Once calibrated, the calibration values remain until you calibrate
again. You can track system calibration and verification results
through the Calibration Trend Report and the System Control Trend
Report.

You must warm up the system before calibrating to stabilize the
system temperatures.