Luminex xyp instrument problems – Luminex 100 IS Version 2.2 User Manual

Page 201

Advertising
background image

x

MAP Technology

Troubleshooting

PN 89-00002-00-069 Rev. A

8 - 11

•

Remove and sonicate or flush the Luminex XYP instrument
Sample Probe to remove possible clogs.

•

Verify that the Sample Probe is properly aligned.

•

Complete the following analyzer functions from the Maintenance
Tab: Backflush, Drain, Alcohol Flush and Water Wash.

•

Verify that the delta cal temperature is

±

3.0 degrees. Recalibrate

if outside these limits and retry. Run controls if calibrating.

•

Power off the system (Luminex 100 analyzer, Luminex XYP
instrument, and PC). Power on the system.

•

Calibrate and verify the system.

Luminex XYP instrument
Problems

Power On

Indicator Light Is

Not On

•

Verify that the Luminex XYP instrument is powered on.

•

Verify that the system is plugged into an outlet with current.

•

If using a surge protector, verify that the power strip is switched
to the On position.

•

Remove the cable and reconnect.

•

Verify the fuses (see “Fuses” on page 7-11).

Heater Block Will

Not Reach Desired

Temperature

•

Verify that the Luminex XYP instrument door is completely
closed. Contact Luminex Customer Support.

There Is a

Grinding Noise
When the Plate

Carrier Moves

•

Verify that the red shipping pin is removed from the Luminex
XYP instrument.

•

Verify that the alignment guide is installed correctly. Contact
Luminex Customer Support.