x

MAP Technology

Using Luminex 100 IS 2.2 Software

PN 89-00002-00-069 Rev. A

6 - 25

the command process in yellow text below the graphic. For example,
Figure 36 shows the Warmup command graphic.

Figure 36. Processing Device Activity Commands

Table 3 shows a recommended schedule for maintenance operations.

Table 3. Maintenance Operations:

Recommended Use Schedule

Operation

Recommended Use Schedule

Warmup

•

Daily

•

After four hours of system inactivity

Prime

•

Daily

•

To remove air from sheath fluid tubing

•

After performing these actions:
– refilling the sheath container
– removing and replacing sheath container
– changing the sheath fluid filter
– changing the syringe seal

Backflush

•

Troubleshooting purposes only:
– to remove obstructions from the cuvette
– if fluid does not flow through the waste

tubing during prime cycles or during
sample acquisition

– if fluid drips from the Sample Probe during

priming and forms puddles of fluid on the
plate

Alcohol Flush

•

Daily

•

After changing the Sample Probe

•

To remove air bubbles from the cuvette using
70% isopropanol