0 functions, 1 parts counting, Functions – Nova-Tech PGL Series User Manual

Page 15: Parts counting

Advertising
background image

7.0  FUNCTIONS

When  weighing,  the  user  can  access  the  applications  that  have  been  enabled  (see 
section 12.2). 

• Weighing 
• Parts counting  
• Percent weighing 
• Check weighing 
• Density determination 
 

The selectable functions can be enabled using a similar method to the Units above 
by turning the functions on or off. 

7.1 

PARTS COUNTING 

This allows the user to weigh a sample of parts to compute an average unit weight 
and then determine the number of items being weighed by dividing the net weight 
by the unit weight value.  The result is always a whole number of parts. 

The balance has a preset number of parts to be used as a sample. These values are 
10, 25, 50 or 100 items. 

Steps: 

1) Press  [Mode]  and  then  the  [Up]  or  [Down]  key  to  select  parts  counting,

“PARTS” will be displayed 

2) Enter parts counting by pressing [Enter] 

3) Press the [Up] or [Down] key to select the sample size, “REF QTY”,  10, 25, 50, 

100, etc., then press [Enter] to confirm 

4) When “LOAD XX Pcs” is shown place XX number of items on the pan and press 

[Enter]  to  compute  the  average  piece  weight.  Display  will  indicate  the  busy 
symbol  

5) Remove the sample when display shows “XX Pcs” and then place an unknown 

quantity  on  the  pan.  The  balance  will  then  compute  the  number  of  parts 
based upon the average piece weight. The display will show the result in Pcs 

6) To count another item press [Mode] and continue from step 2. 

15 |

P a g e

© Adam Equipment Company 2008

7) To return to normal weighing, press [Mode] then press [Esc] 

Advertising