8.0  CALIBRATION

 

The PGL series can only be calibrated with an external mass. 

8.1 

MANUAL CALIBRATION 

Pressing the [Cal] key will start calibration. Calibration can also be called for after a 
set time period as determined by the user (see section 12.5). 

Calibration using External Calibration mass: 

1) Pressing [Esc] will abort the calibration at any time 

2) Check the display is at zero. Tare if necessary 

3) Press the [Cal] key 

4) The  display  will  show  the  balance  setting  a  new  Zero  condition  by  showing 

“LOAD 0”. Make sure the pan is empty then press the [Enter] key to continue 

5) The  balance  will  then  show  the  value  of  the  calibration  mass  required 

sounding a beep, for example “LOAD 1000 g”  

6) If  a  different  mass  is  desired  press  the  up  or  down  arrow  key  to  show  the 

alternate masses possible for the scale.   

 

PGL 

203 

PGL 

303 

PGL 

2002 

PGL 

3002 

PGL 

4001 

PGL 

6001 

PGL 

8001 

PGL 

10001 

PGL 

20001 

Mass1 

100 g 

100 g 

1000 g 

1000 g 

2000 g 

3000 g 

4000 g 

5 kg 

10 kg 

Mass2 

200 g 

200 g 

2000 g 

2000 g 

3000 g 

4000 g 

5000 g 

10 kg 

20 kg 

Mass3 

 

300 g 

 

3000 g 

4000 g 

5000 g 

6000 g 

 

 

Mass4 

 

 

 

 

 

6000 g 

7000 g 

 

 

Mass5 

 

 

 

 

 

 

8000 g 

 

 

22 |