12.2  ENABLE WEIGHING MODES

Same steps are followed to enable or disable the weighing modes. 

1) Press [Enter] when “MODES” is displayed. The display will show the first mode 

i.e., Parts Counting (“PARTS”) together with its enabled state “OFF” or “On”. 
The  user  can  enable  or  disable  the  parts  counting  mode  by  using  [Up]  or 
[Down]
. Pressing [Enter] will confirm the setting and will advance to the next 
weighing mode. Repeat for each mode in turn 

2) Press [Mode] to advance to setting of the next menu or press [Esc] to return 

to normal weighing 

12.3  ENABLE SERIAL INTERFACE PARAMETERS 

The parameters affecting the serial setup are set in a similar manner to the other 
parameters. 

Press [Enter] when “SERIAL” is displayed to enter the sub‐menu.  

The parameters that can be set are:   

ENABLE 

The serial port can be set to On or OFF 

BAUD RATE 

Set the Baud Rate to 4800, 9600, 19200 or 38400 

PARITY 

Set the Parity to NONE, EVEN or ODD 

8 data bits and one stop bit are always used, (8n1, 8E1, 8o1). 

STABLE 

To print when stable (On) or regardless of stability (OFF) 

CONTINUOUS 

Set the RS‐232 to send data continuously to On or OFF 

PERIODIC 

Set the RS‐232 to send data periodically (set in seconds) to On or OFF. If 
On  is  selected,  the  value  can  be  changed  between  1  and  999  seconds, 
using [Up] and [Down]  

FORMAT 

 

To send data as a single line of data, using the standard format or using a 
customer designed format (FORM 1 or FORM 2).  

During  continuous  or  periodic  printing  only  the  weight  is  printed.  Forms 
are not available. 

31 |

P a g e

© Adam Equipment Company 2008