Dissolved oxygen / atc electrode – Nova-Tech DO 700 User Manual

Page 10

Advertising
background image

3.

DISSOLVED OXYGEN / ATC ELECTRODE

The DO 700 includes a galvanic DO
electrode with built in temperature
measurement.

The electrode has a dual cable; a BNC
connector for DO measurement and a mini-
phono plug for temperature. The electrode is
pre-filled with electrolyte that is separated
from the sample by an oxygen permeable
membrane. The membrane is pre-assembled
for you and fixed to a detachable cap. The
pre-assembled cap design allows simple
replacement and fast conditioning.

The electrode is light-weight with a 12-mm
body diameter and a 16-mm tip diameter
from at the detachable Noryl cap. The
compact sensing area reduces air
entrapment resulting in quick, accurate, and
stable readings.

Provide simple stirring for best results—
ideally achieve a minimum water flow rate of
2 inch/second on the membrane.

The probe is not recommended for samples
beyond 0 to 50 °C.

The membrane is thin and can be replaced if damaged. Use care to
protect from scratches, abrasion, or contact with solids. For best
results keep membrane clean by rinsing after daily use. See

Section 7

– Electrode Maintenance

.