5 mg/l or ppm (concentration) calibration, Mg/l or ppm (concentration) calibration – Nova-Tech DO 700 User Manual

Page 14

Advertising
background image

4) Press

. The CAL indicator will blink and the secondary display

will indicate the calibration point, 0.0.

5) Press

to confirm the calibration. The primary display will blink

briefly before calibrating to 0.0%.

6) Press

to return to measurement mode, or continue with 100%

calibration if desired (note 0% calibration followed by 100%
calibration is not recommended---always perform 100% calibration
first).

4.5 mg/L or ppm (Concentration) Calibration

Performing a saturation calibration at 100% or 0% will simultaneously
calibrate the corresponding mg/L (ppm) concentration value.
Therefore, additional mg/L (ppm) calibration isn’t required in most
circumstances.

If desired, you can perform a calibration adjustment in mg/L (ppm)
without affecting your % saturation calibration value.

1) Rinse the probe well with clean water (do not dry).

2) Press to select concentration mg/L (ppm).

3) Dip the probe into a sample of known oxygen concentration (i.e.

determined by titration or another instrument). Wait for the reading
to stabilize.

4) Press

. The CAL indicator and current concentration display

briefly, before CAL flashes. The concentration value based on the
factory default calibration is shown.

5) Press

or

to adjust the reading to match the known

oxygen concentration value.

6) Press

to confirm the calibration. The primary display will blink

briefly before returning to measurement mode.