7 3.4 salt (salinity adjustment), 8 3.5 ofs (offset adjustment) from % mode, 4 salt (salinity adjustment) – Nova-Tech DO 700 User Manual

Page 18: 5 ofs (offset adjustment) from % mode, 5 unit (unit adjustment) from concentration mode, Section 5.8 – offset adjustment

Advertising
background image

5.7 3.4 SALt (Salinity Adjustment)

For best accuracy, enter the actual salinity value if your samples are at
least 0.1 ppt (100 ppm). Maximum adjustment is 50 ppt (50,000 ppm or
5% salt).

Press

.

Press

or

to adjust the value (in parts per thousand).

Tip: 0.1 ppt = 100 ppm salinity

Press

to confirm.

5.8 3.5 OFS (Offset Adjustment) from % mode

Use this feature to offset the meter’s value to another DO meter.
Using the % saturation mode, observe the reading a sample solution
after it has stabilized. Similarly, observe the reading of the same
sample using another DO meter as a reference. The probe of the
reference meter should be immersed in the same sample at the same
depth.

The DO 700 allows +/- 10.0% offset adjustment. Note that subsequent
user calibrations will reset the offset adjustment back to the factory
default value of 0.0 %.

Press

.

Press

or

to adjust the % offset.

Press

to confirm.

5.9 3.5 Unit (Unit Adjustment) from concentration mode

Press

.

Press

or

to select ppm or mg/L units.

Press

to confirm.

12

Advertising