Troubleshooting guide – Nova-Tech DO 700 User Manual

Page 30

Advertising
background image

10.

TROUBLESHOOTING GUIDE

PROBLEM CAUSE

SOLUTION

No display

Main power not switched
on. AC Adapter socket not
inserted properly.

Switch on the power supply.

Re-insert AC Adapter.

“Ur” (Under
range)

“Or” (Over
range)

in primary or
secondary
display

DO or Temperature value
is out of range.

Probe is not connected.

Electrode dirty or broken.

Meter not calibrated.

Confirm measurement condition,
ensure temp is within range.

Check electrode connections.

Check electrode-add electrolyte
&/or replace membrane.

Reset meter & recalibrate.

Meter not
responding to
key press

Manual HOLD or Auto
HOLD is active.



Worn keypad.

Press

to deactivate manual

HOLD.

See

Section 5.4 — Auto Hold

to

disable Auto Hold feature if
enabled.

Contact Technical Service.

Invalid key; Button is not
functional in the current
operation mode.

Press alternate key.
Select valid key depending on
current mode.

Secondary
display
continually
shows
0 and 100,
during %
calibration

The calibration value is too
high to be accepted as a
0% standard, and too low
to be accepted as a 100%
standard (10.1% to
49.9%), compared against
factory default value.

Service probe.
Check calibration solution.
Check sample conditions (temp,
salinity and pressure settings).
Reset meter.