Setup functions, 1 1.0 cal (calibration), 2 2.0 ele (electrode information) – Nova-Tech DO 700 User Manual

Page 16: 0 cal (calibration), 0 ele (electrode information)

Advertising
background image

5.

SETUP FUNCTIONS

Use the setup feature to customize your instrument operation.

There are two setup menus; one menu is derived from the %
saturation mode and the other from the mg/L (or ppm) concentration
mode.

First, select the appropriate measurement mode you wish to adjust by

pressing

until the desired units are displayed. During

measurement, press and hold

for 5 seconds to enter SETUP

mode of the parameter being measured.

Press

or

to change programs or change options.

Press

to select the program or confirm selection.

Press

to go back one level or return to measurement mode.

5.1 1.0 CAL (Calibration)

Press

to view stored calibration points.

Note: % saturation mode will list two points, while
concentration mode has only one point.

5.2 2.0 ELE (Electrode Information)

Press

to view calibration factor (CAL.F)

TIP: The factor is an indication of the probe’s efficiency. It is
the ratio of the actual mV produced by the probe to the
theoretical mV value. The ratio displays from 0.5 to 1.999.

Press

to view mV output for 100% (HI).

Press

to view mV output for 0% (LO).

10