PROBLEM CAUSE

SOLUTION

CAL

% calibration value is not
acceptable.
Electrode is disconnected
or failing.

Check electrode connections.
Check that correct DO unit is
selected & repeat cal procedure.
Service electrode.
Reset meter.
Replace electrode.

CAL

ppm or mg/L calibration is
attempted when the factory
calibrated absolute value is
below 2.00

Use concentration calibration
solution higher than 2.00.
Check electrode connections.
Check that correct DO unit is
selected & repeat cal procedure.
Service electrode.
Reset meter.
Replace electrode.

Slow
response or
unstable
readings

External noise pickup or
induction caused by
nearby electric motor

Insufficient electrolyte in
probe

Air bubbles near sensor


Dirty or damaged probe

Probe not deep enough in
sample

Broken probe

Remove or switch off interfering
device

Fill probe with electrolyte & / or
replace pre-assembled cap

Stir or tap probe to remove
bubbles, reduce stirring

Clean the probe and re-calibrate

Submerse sensor cap into sample


Replace probe or membrane