Taking measurements – Nova-Tech DO 700 User Manual

Page 20

Advertising
background image

6.

TAKING MEASUREMENTS

During measurement the dissolved oxygen reading is
automatically compensated for pressure and salinity. For best
accuracy, ensure the pressure and salinity settings are adjusted
accordingly from the setup menu. The factory default value for
barometric pressure is 760 mm Hg or 101.3 bar (sea level). The
factory default value for salinity is 0.0 ppt (no salinity).

See

Section 5.6 — Barometric Pressure Adjustment

.

See

Section 5.7 — Salinity Adjustment

.

Press

to

select the desired measurement mode:

 Percentage Saturation (%) and Temp (

º

C or

º

F)

 Concentration (mg/L or ppm) and Temp (

º

C or

º

F)

Follow these general rules when taking measurements:

 Keep the membrane free from contact with solid

objects.

 Provide stirring of your solutionthis helps to overcome

oxygen consumption of the probe and prevents air
bubble entrapment.

 Do not submerge the cable for extended periods.
 Use with aqueous solutions only.
 The pre-assembled cap must be completely submersed

to obtain an accurate reading in solutions.

 Rinse the probe well with DI water or rinse solution

between measurements.