take readings:

1.

Rinse the probe with deionized or

distilled water before use

to r
emove any

impurities adhering to the pr

obe body

.

If the pH electr

ode has dehydrated, soak

it for 30 minutes in OAKT

ON electr

ode

storage solution or a 2M–4M KCl

solution.

2.

Press ON to switch on meter

.

The

MEAS

annunciator appears on the top

center of the LCD. The

A

TC indicator

appears in the lower right hand corner

to indicate

Automatic T

emperatur

e

Compensation (See page 19 to set

Manual T

emperatur

e Compensation).

See figur

e

3.

Dip the probe into the sample.

When dipping the pr

obe into the sample,

the sensor or the glass bulb of the electr

ode

must be completely immersed into the

sample. Stir the pr

obe gently in the sample

to cr

eate a homogenous sample.

4.

Allow time for the reading to stabilize.

Note the r

eading on the display

.

5.

T

o

toggle between pH and mV (or Rel

mV) r

eadings,

press the MODE key

.

21

20

Taking Measurements

Be sur

e to r

emove the electr

ode soaker bottle or pr

otective r

ubber cap on the

electr

ode befor

e measur

ement.

pH

ATC

MEAS

C

A

A

5.3

T

aking measurements with READY indicator selected on

If the READY

indicator has been activated, the

READY

annunciator lights when

the r

eading is stable*. Switch the READY

indicator on or of

f in Set up pr

ogram

P4.1—see page 32 for dir

ections.

*The READY

indicator appears and the r

eading holds until the measur

ed value exceeds the tolerance


0.02 pH; ±

0.8 mV <400; ±

1.2 mV >400). Then, the

READY

annunciator turns of

f.

T

aking measurements with the auto endpoint feature selected on

NOTE:

this featur

e is available on model pH 310 only

.

When a r

eading is stable for mor

e than 5 seconds, the auto endpoint featur

e will

automatically “hold” the r

eading. The “hold” indicator appears on the left side of

the display

. Pr
ess the HOLD key to r

elease the r

eading. Switch the

Auto endpoint

featur

e on or of

f in Set up pr

ogram P4.1—see page 32 for dir

ections.