Error messages, T roubleshooting – Nova-Tech pH 310 User Manual

Page 22

Advertising
background image

43

42

11. Error Messages

LCD

Indicates

Cause

Solution

Display

Err

Unr

ecognized input

W

rong input in

Release key

. Select

annunciator

fr

om keypad

selected mode.

valid operations

depending on mode.

CAL

&
Err

Calibration err

or

W

rong buf

fer value

Check your buf

fer input

annunciators on/

input at calibration.

value, clean pr

obe.

Buf

fer and

Dirty pr

obe.

See Calibration sections or

electrode

Pr

obe Maintenance section.

indicators blink

Battery

Low battery level

Need new batteries

Clean battery contacts.

indicator blinks

or battery

Replace batteries with fr

esh

connection is bad

ones, noting polarity

* See "W

arranty" and "Return of Items" on page 47

If an err

or message appears in the primary

display (the upper r

ow of lar

ger digits),

switching of

f the meter and switching it on

again may eliminate the err

or message.

See figur

e

If err

or persists, or the meter shows

incorr

ect values, r

eturn the meter

.

For a complete diagram of the display

,

see page 5

.

ERR 1 in primar

y display

A

A

10. T

roubleshooting

Problem

Cause

Solution

Power on but

a) Batteries not in place.

a) Check that batteries ar

e in

no display

place and making good

contact.

b) Batteries not in corr

ect

b) Reinsert batteries with

polarity (+ and –).

corr

ect polarity

.

c) W

eak batteries.

c) Replace batteries or attach

optional AC

adapter

.

Unstable

a)
Air bubbles in pr

obe.

a) T

ap pr

obe to r

emove bubbles.

readings

b) Dirty pr

obe.

b) Clean the pr

obe and

recalibrate.

c) Pr

obe not deep

c) Make sur

e sample entir

ely

enough in sample.

covers the pr

obe sensors.

d) External noise pickup

d) Move or switch of

f

or induction caused by

interfering motor

.

nearby electric motor

.

e) Br

oken pr

obe.

e) Replace pr

obe. See page 45.

Slow response

a) Dirty/Oily pr

obe.

a) Clean pr

obe. See "Pr

obe

Car

e & Maintenance",

page 40.