Ph calibration – Nova-Tech pH 310 User Manual

Page 8

Advertising
background image

1.

If necessary

, press the MODE key to

select pH mode

. The pH indicator

appears in the upper right hand corner

of the display

.

See figur

e

2.

Rinse the probe thoroughly with

de-ionized water or a rinse solution.

Do

not wipe the pr

obe; this causes a build-up

of electr

ostatic char

ge on the glass surface.

3.

Dip the probe into the calibration buf

fer
.

The end of the pr

obe must be completely

immersed into the sample. Stir the pr

obe

gently to cr

eate a homogeneous sample.

4.

Press CAL/MEAS

to enter pH calibration

mode. The CAL

indicator lights. The

primary display will show the measur

ed

reading while the smaller secondary dis-

play will indicate the pH standar

d buf

fer

solution.

See figur

e

5.

W

ait for the measured pH value to

stabilize.

See figur

e

15

14

6.

Press ENTER to confirm calibration

.

The meter is now calibrated to the

curr

ent buf

fer
.

See figur

e

• If you ar

e performing multipoint

calibration, go to step 7.

• If you ar

e performing one-point

calibration, go to step 9.

7.

Rinse the probe with de-ionized water

or a rinse solution, and place it in the

next pH buf

fer
.

8.

Follow steps 5 to 8 for additional

calibration points.

9.

When calibration is complete,

press CAL/MEAS to return to pH

measurement mode

.

Notes

T

o

exit from pH Calibration mode without

confirming calibration,

DO NOT pr

ess

ENTER

in step 6. Pr

ess
CAL/MEAS

instead.

If the selected buf

fer value is not within

±1.0 pH fr

om the measur

ed pH value: the

electr

ode and buf

fer icon blink and the ERR

annunciator appears in the lower left cor-

ner of the display

.

T

o

limit the number of pH buf

fer values

available during calibration, see section

P4.2 on page 33.

pH

CAL

pH

READY

D

C

D

B

C

pH

CAL

pH

pH

CAL

pH

READY

B

pH

ATC

MEAS

C

A

A

pH calibration

NOTE:

W

e r

ecommend that you perform at least a 2-point calibration using standar

d

buf
fers that bracket (one above and one below) the expected sample range.

Y

ou can perform a 1-point calibration, but make sur

e that the buf

fer value is

close to the sample value you ar

e measuring.

4.3

C

If the READY

indicator has been activiated

(set up program P4.1—see page 32), the READY

annunciator lights when the reading is stable.

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(86