P5.0: resetting to factor y default settings, P5.0: setting the real-time clock – Nova-Tech pH 310 User Manual

Page 19

Advertising
background image

37

36

P5.0: Resetting to factor

y default settings

A

vailable in pH 300 meter only

In the pH 300 meter

, Pr
ogram 5 lets you r

eset all parameters to factory default

settings. This clears all calibration data, memory

, and any other setup functions you

might have changed.

MEAS

SETUP

A

MEAS

SETUP

B

8.6

From measurement mode

1.

Press Setup key

to enter Set Up mode.

2.

Press the

▲
and

▼
keys

to scr

oll thr

ough

subgr

oups until you view parameter

P5.0.

See figur

e

3.

Press the ENTER

key

to enter

parameter P5.0.

See figur

e

4.

Press the

▲
and

▼
keys

to toggle

between NO

and

YES.

• NO r

etains curr

ent settings

•

YES

resets to factory default settings

5.

Press the ENTER

key

to confirm selec-

tion and to r

eturn to the subgr

oup menu.

Pr

ess the CAL/MEAS key to r

eturn to

measur

ement mode.

A

B

P5.0: Setting the real-time clock

A

vailable in pH 310 meter only

The pH 310 meter featur

es a r

eal-time calendar and clock. This helps you meet GLP

(Good Laboratory Practice) standar

ds.

8.7

SETUP

A

SETUP

B

From measurement mode

1.

Press Setup key

to enter Set Up mode.

2.

Press the

▲
and

▼
keys

to scr

oll thr

ough

subgr

oups until you view parameter

P5.0.

See figur

e

3.

Press the ENTER

key

to enter

parameter P5.0. The meter lets you

select the century: “19-” or “20-”.

The century digits will flash.

See figur

e

4.

Press the

▲
and

▼
keys

to toggle to the

corr

ect century

.

5.

Press the ENTER

key

to confirm the

century and move to “year” selection.

The “year” digits will flash.

See figur

e

6.

Press the

▲
and

▼
keys

to toggle to the

corr

ect year

.

7.

Press the ENTER

key

to confirm the

year and move to “month” selection.

The “month” digits will flash.

See figur

e

8.

Press the

▲
and

▼
keys

to toggle to the

corr

ect month.

9.

Press the ENTER

key

to confirm the

month and move to “date” selection.

The “date” digits will flash.

See figur

e

10.

Press the

▲
and

▼
keys

to toggle to

the corr

ect date.

A

B

SETUP

C

C

D

SETUP

D

SETUP

E

E

continued on next page