P4.0: unit configuration – Nova-Tech pH 310 User Manual

Page 17

Advertising
background image

From measurement mode

1.

Press Setup key

to enter Set Up mode.

2.

Press the

▲
and

▼
keys

to scr

oll thr

ough

subgr

oups until you view parameter P4.0.

See figur

e

3.

Press the ENTER

key

twice to select

parameter 4.2.

See figur

e

4.

Press the

▲
and

▼
keys

to select 2, 3, 4,

or 5 point pH calibration.

5.

Press the ENTER

key

to confirm selection

and to r

eturn to the subgr

oup menu.

Pr

ess the CAL/MEAS key to r

eturn to

measur

ement mode.

From measurement mode

1.

Press Setup key

to enter Set Up mode.

2.

Press the

▲
and

▼
keys

to scr

oll thr

ough

subgr

oups until you view parameter P4.0.

See figur

e

3.

Press the ENTER

key

to select

parameter 4.1.

See figur

e

4.

Press the

▲
and

▼
keys

to select the

configuration you r

equir

e.

•OFF switches the READY

indicator of

f.

•ON switches the READY

indicator on.

•

pH 310 meter only:

ON and HOLD

together switches the auto endpoint

featur

e on.

5.

Press the ENTER

key

to confirm selection

and to pr

oceed to step 4 of P

.4.2. Pr

ess the

CAL/MEAS key to r

eturn to measur

ement

mode.

Notes

Meter default is set for Ready indicator on,

and auto endpoint function of

f.

33

32

P4.0: Unit configuration

P4.1: READY

indicator and auto endpoint function

Pr

ogram P4.1 lets you select “READY

indicator on” to indicate when the r

eading is

stable*, or select “READY

indicator of

f” for faster meter r

esponse.

pH 310 meter only:

Pr

ogram P4.1 also lets you switch the

Auto endpoint function on

or of

f. Select auto endpoint on to “hold” the r

eading when it is stable for mor

e than

5 seconds. The display automatically fr

eezes, and the HOLD

indicator appears on

the left side of the display

. Pr
ess the HOLD key to r

elease the display and access

other functions. Select auto endpoint of

f to deactivate this featur

e.

SETUP

READY

SETUP

ON

8.5

A

A

B

B

SETUP

READY

SETUP

ON

C

C

D

D

P4.2: Selecting number of pH calibration points

Pr

ogram P4.2 lets you select the number of calibration points that appear in pH

calibration mode: 2, 3, 4, or 5. This lets the meter scr

oll thr

ough the calibration points

mor

e quickly if you r

egularly calibrate at less than 5 points.

*The READY

indicator appears and the r

eading holds until the measur

ed value exceeds the tolerance

(±
0.02 pH; ±

0.8 mV <400; ±

1.2 mV >400). Then, the

READY

annunciator turns of

f.