Temperature calibration, Relative mv calibration – Nova-Tech pH 310 User Manual

Page 9

Advertising
background image

T

emperature calibration

1.

Make sur

e the

A

TC pr

obe (or

temperatur

e connector of the “All-in-

One” electr

ode) is attached to the 6-pin

connector

. The

A

TC annunciator will

appear at the right-hand side of the LCD.

2.

Switch the meter on.

Press the MODE

key

to select pH mode.

3.

Press the CAL/MEAS key

to enter pH

calibration mode. The CAL

indicator will

appear above the primary display

.

4.

While in pH calibration mode,

press the

MODE

key

to enter temperatur

e calibra-

tion mode. The primary display shows the

temperatur

e value with the last set of

fset

and the secondary display shows the

factory default temperatur

e value.

See figur

e

5.

Dip the A

TC probe

(or “All-in-One”

electr

ode)

into a solution of known

temperature

(i.e. a temperatur

e bath).

Allow time for the temperatur

e pr

obe to

stabilize.

6.

Scroll with the


and


keys

to set

the corr

ect temperatur

e value (i.e. the

temperatur

e of the temperatur

e bath).

Y

ou can adjust the r

eading in incr

ements

of 0.1°C.

See figur

e

7.

Once you have selected the corr

ect

temperatur

e,

press the ENTER key

.

The meter automatically r

eturns to

pH measur

ement mode.

17

16

Temperature Calibration

The temperatur

e sensor is factory calibrated. Calibrate your sensor only if you sus-

pect temperatur

e err

ors that may have occurr

ed over a long period of time or if you

have a r

eplacement pr

obe.

4.5

ATC

CAL

C

B

B

ATC

CAL

C

A

A

Relative mV Calibration

1.

While in the measurement function,

press MODE

to enter the mV mode.

The mV indicator appears in the upper

right hand corner

.

See figur

e

2.

Press the CAL/MEAS key

. The

CAL

indicator appears above the primary

display

. The primary display shows

the

relative mV r

eading and the secondary

display shows the absolute mV value.

NOTE: If you have never calibrated

relative mV or if the meter has been r

eset,

the value shown in the primary display is

the same as the absolute mV value.

See figur

e

3.

Press the

or

keys

to enter the

relative mV value that matches your

desir

ed r

eading.

See figur

e

4.

Press the ENTER key

to confirm

the r

eading and t

o

r

eturn to the

measur

ement mode. The primary

display now shows the r

elative mV

reading. The RmV indicator appears in

the upper right hand corner

.

Notes

T

o

view the mV of

fset value, use the

SETUP

mode Pr

ogram P3.1. See page

30 for instr

uctions.

The r

elative mV (RmV) indicator appears

whenever the mV of

fset is not zer

o.

T

o

r

eset the all calibration and of

fset values

in memory to the factory default settings,

use:

pH 300 meter

:

SETUP

Pr

ogram P

5.0.

See page 36.

pH 310 meter:

SETUP

Pr

ogram P6.0.

See page 39.

4.4

CAL

mV

C

C

CAL

mV

B

B

ATC

MEAS

mV

C

A

A

Notes

Y

ou can of

fset the temperatur

e r
eading up to ±

5°C fr

om the original r

eading.

T

o

exit this pr

ogram without confirming the temperatur

e calibration value, DO NOT

pr

ess
ENTER

in step 7. Pr

ess
CAL/MEAS

instead.