Range Mode
AutoRange is preset into the system as the default. AutoRanging lets the
system select the range based on current oxygen levels. Manual range allows
you to scroll through the ranges and set your own desired range value.
AutoRange Function
1. After completing initial start-up, you
may select
RANGE
from the Analyze
mode main menu.
2. After pressing
RANGE
, the screen
directs you to select from auto or fixed
(manual) range. Press
AUTO
to select
AutoRange.
3. The next screen returns you to the
Analyze mode menu. The last line
indicates the range selection and mode.
VALVES: ANALYZE
T=25C
C=9:59
FLO W = 150.0 CC/M IN
M AN RANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
SETUP
STDBY
CAL
VALVES: ANALYZE
T=25C
C=9:59
SELECT AUTO-RANGE OR
FIXED-RANGE FO R O2.
1
2
3
4
5
AUTO
FIXED
ESC
VALVES: ANALYZE
T=25C
C=9:59
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
SETUP
STDBY
CAL